Ph Controller Dosing Manufacturer – SUP-PTU300 Turbidity meter – Sinomeasure
Ph Controller Dosing Manufacturer – SUP-PTU300 Turbidity meter – Sinomeasure Detail:
-
Specification
| Range | 0-20 NTU (31),0-1 NTU (30) |
| Operating voltage | DC 24V |
| Measurement | 90° scattering |
| Working mode | Continuous monitoring of drainage, intermittent automatic discharge |
| Zero drift | ≤±0.015 NTU |
| Value error | ≤±2% or ±0.015 NTU larger |
| Discharge mode | Automatic discharge |
| Calibration | Formalhydrazine standard liquid calibration (factory calibrated) |
| Water pressure | 0.1 Kg/cm3-8Kg/cm3, flow not exceeding 300 mL /min |
| Digital output | RS485Modbus protocol (baud rate 9600,8, N 、1) |
| Analog output | 4-20 mA |
| Storage temperature | -20℃-60℃ |
| Working temperature | 0-50℃ |
| Sensor material | Composite |
| Maintenance cycle | 6-12 months recommended (depending on site water quality environment) |
-
Introduction




Product detail pictures:
Related Product Guide:
We rely upon strategic thinking, constant modernisation in all segments, technological advances and of course upon our employees that directly participate in our success for Ph Controller Dosing Manufacturer – SUP-PTU300 Turbidity meter – Sinomeasure , The product will supply to all over the world, such as: Romania, Paraguay, Istanbul, As a well educated, innovative and energetic staff,we are responsible for all elements of the research, design, manufacturing, sales and distribution. With studying and developing new techniques,we are not only following but also leading the fashion industry. We listen attentively to the feedback from our customers and provide instant communication. You will instantly feel our expertise and attentive service.
In China, we have many partners, this company is the most satisfying to us, reliable quality and good credit, it is worth appreciation.