Specification
| Range | 0-20 NTU (31),0-1 NTU (30) |
| Operating voltage | DC 24V |
| Measurement | 90° scattering |
| Working mode | Continuous monitoring of drainage, intermittent automatic discharge |
| Zero drift | ≤±0.015 NTU |
| Value error | ≤±2% or ±0.015 NTU larger |
| Discharge mode | Automatic discharge |
| Calibration | Formalhydrazine standard liquid calibration (factory calibrated) |
| Water pressure | 0.1 Kg/cm3-8Kg/cm3, flow not exceeding 300 mL /min |
| Digital output | RS485Modbus protocol (baud rate 9600,8, N 、1) |
| Analog output | 4-20 mA |
| Storage temperature | -20℃-60℃ |
| Working temperature | 0-50℃ |
| Sensor material | Composite |
| Maintenance cycle | 6-12 months recommended (depending on site water quality environment) |
-
Introduction




Product detail pictures:
Related Product Guide:
Fast and good quotations, informed advisers to help you choose the correct product that suits all your needs, a short production time, responsible quality control and different services for paying and shipping affairs for Orp Analyzer Supplier – SUP-PTU300 Turbidity meter – Sinomeasure , The product will supply to all over the world, such as: Russia, Istanbul, Bangalore, Insisting over the high-quality generation line management and prospects guide provider, we've made our resolution to offer our shoppers using the initially stage purchasing and soon after provider working experience. Preserving the prevailing helpful relations with our prospects, we even now innovate our product lists the many time to meet up with the brand new wants and stick to the latest trend of this business in Ahmedabad. We're ready to facial area the difficulties and make the transform to grasp many of the possibilities in international trade.
We are long-term partners, there is no disappointment every time, we hope to maintain this friendship later!