Medidor Ph Agua Manufacturers – SUP-PH8001 Digital pH sensor – Sinomeasure
Medidor Ph Agua Manufacturers – SUP-PH8001 Digital pH sensor – Sinomeasure Detail:
-
Specification
Product | Digital pH sensor |
Model | SUP-PH8001 |
Measurement range | 0.00-14.00pH; ±1000.0mV |
Resolution | 0.01pH,0.1mV |
Heat resistance | 0 ~ 60℃ |
Output | RS485 (MODBUS-RTU) |
ID | 9600,8,1,N (Standard) 1-255 |
Power supply | 12VDC |
Power consumption | 30mA @12VDC |
-
Introduction
-
Communication Protocol
Communication interface: RS485
Port setting: 9600,N,8,1 (default)
Device address: 0×01 (default)
Protocol specification: Modbus RTU
Instruction support: 0×03 read into register
0×06 write register | 0×10 write register continuously
Register data format
Address | Data name | Conversion factor | Status |
0 | Temperature | [0.1℃] | R |
1 | PH | [0.01pH] | R |
2 | PH.mV | [0.1mV] | R |
3 | PH. Zero | [0.1mV] | R |
4 | PH. slope | [0.1%S] | R |
5 | PH. Calibration points | - | R |
6 | System status. 01 | 4*bits 0xFFFF | R |
7 | System status. 02 | 4* bits 0xFFFF | R/W |
8 | User command address | - | R |
9 | User commands. Results | [0.1mV] | R |
11 | ORP | [0.1mV] | R |
Product detail pictures:




Related Product Guide:
With a complete scientific quality management system, good quality and good faith, we win good reputation and occupied this field for Medidor Ph Agua Manufacturers – SUP-PH8001 Digital pH sensor – Sinomeasure , The product will supply to all over the world, such as: Zambia, Adelaide, Danish, We'll supply much better products with diversified designs and expert services. We sincerely welcome friends from over the world to visit our company and cooperate with us on the basis of long-term and mutual benefits.

The customer service staff's answer is very meticulous, the most important is that the product quality is very good, and packaged carefully, shipped quickly!