Buy Mag Flow Suppliers – SUP-825-J Signal Calibrator 0.075% high accuracy – Sinomeasure
Buy Mag Flow Suppliers – SUP-825-J Signal Calibrator 0.075% high accuracy – Sinomeasure Detail:
-
Specification
| General Specifications | Operating temperature | -10℃~55℃ |
| Storage temperature | -20℃~70℃ | |
| Relative Humidity (%RH operating without condensation) | 90%(10℃~30℃) | |
| 75%(30℃~40℃) | ||
| 45%(40℃~50℃) | ||
| 35%(50℃~55℃) | ||
| Uncontrolled<10℃ | ||
| EMC | EN55022, EN55024 | |
| Vibration | Random, 2g, 5 to 500Hz | |
| Concussion | 30g, 11ms, half sine bow wave | |
| Power requirement | 4 AA Ni-MH, Ni-Cd batteries | |
| Size | 215mm×109mm×44.5mm | |
| Weight | About 500g |
| DC Voltage | Range | Accuracy |
| Measurement | (0~100)mVDC(Upper display) | ±0.02% |
| (0~30)VDC(Upper display) | ±0.02% | |
| (0~100)mVDC(Lower display) | ±0.02% | |
| (0~20)VDC(Lower display) | ±0.02% | |
| Source | (0~100)mVDC | ±0.02% |
| (0~10)VDC | ±0.02% |
| Resistance | Range | Accuracy | |
| 4-wire | 2-, 3-wire | ||
| Accuracy | Accuracy | ||
| Measurement | (0~400)Ω | ±0.1Ω | ±0.15Ω |
| (0.4~1.5)kΩ | ±0.5Ω | ±1.0Ω | |
| (1.5~3.2)kΩ | ±1.0Ω | ±1.5Ω | |
| Excitation Current:0.5mA
Clear of resistance before measuring according to ‘10.4 Clear of Resistance and RTDs’. *3-wire: Assumes matched leads with a total resistance not exceeding 100Ω. Resolution (0~1000)Ω: 0.01Ω; (1.0~3.2)kΩ: 0.1Ω. |
|||
-
Advantages
· Two separate channels display.
the upper display shows measure parameters ;
the lower one shows measure or source parameters;
· Counting pulse function
· Calibration functions
· Auto ramping and auto stepping
· Manual and automatic cold junction compensating
· Clear function
· Temperature unit switching
· Auto flashing jacks
· Backlight LCD
· Battery gauge
Product detail pictures:
Related Product Guide:
It adheres on the tenet "Honest, industrious, enterprising, innovative" to acquire new solutions continuously. It regards prospects, success as its personal success. Let us build prosperous future hand in hand for Buy Mag Flow Suppliers – SUP-825-J Signal Calibrator 0.075% high accuracy – Sinomeasure , The product will supply to all over the world, such as: Iraq, Estonia, Roman, We always insist on the management tenet of "Quality is First, Technology is Basis, Honesty and Innovation".We are able to develop new products continuously to a higher level to satisfy different needs of customers.
The factory workers have rich industry knowledge and operational experience, we learned a lot in working with them,we are extremely grateful that we can encount a good company has excellent wokers.